Mitech MHVS-1digital micro hardness tester, based on the principle that positive quadrangular pyramid diamond indenter presses the surface of the sample to produce indentation.By measuring the diagonal length of the indentation to achieve the measurement of the hardness of the material can be for small specimens, thin specimens, surface coating, heat treatment of the workpiece surface Vickers hardness test. Adopted computer-aided, high-rate optical measurement system, photoelectric sensor technology, adjustable light, it is the upgrading products of universal type of micro hardness tester and widely used in the fields of metal processing, electronics industry, mold parts, watch manufacturing, engineering quality inspection and so on. It is an ideal hardness tester for material research and testing.
Micro-Vickers (or Knoop) hardness test principle is that put the provisions of the positive pyramid diamond indenter into the sample surface(with fixed experimental force) and maintain a certain length (holding), and then unloading. Finally, there is a positive quadrangular pyramid or kenup indentation with a square surface on the surface of the specimen. Then we can attain the area of indentation via measuring the length of the diagonal by a micrometer eyepiece. Then the corresponding Vickers (or Knoop) hardness values are obtained.
Usually Vickers hardness values can be converted according to the following formula
HV=constant×test force / indentation surface area≈0.1891 F/d2
Note:
HV, Vickers hardness symbols
F, test force
d, the arithmetic mean of of the two diagonal d1, d2
Operation Temperature:18~28℃;
Relative Humidity:≤65%;
In an environment free from vibration,no corrosive medium in surrounding.
Installed on a flat basis.
<!--[if !supportLists]--><!--[endif]-->
Read the manual carefully before using the instrument. Learn the operation steps and attentions to avoid damage the instrument or personal safety accidents caused by improper operation;
Carefully remove the shockproof tape for protecting the indenter when installed otherwise the accuracy of indenter positioning is not high.
Machine add unloading test power, it is forbidden to turn the variable load the handwheel, such as rotating dislocation can lead to internal gear, chaos test force, and even cause damage to items within the machine;
Don’t apply the test force when measuring. If not carefully press the buttonto start the test force to load, be sure to wait for the test force after the completion of the next step.
Diamond indenter and indenter are the core of the instrument, the operation should be careful not to touch the pressure head
The indenter should be kept clean when used to ensure the accuracy of the test. If the indenter is stained with oil or dust and other stains, can be used cotton with absorbent (industrial) or ether in the top of the indenter carefully wipe clean
When not in use for a long time should cover dust cover, in order to prevent dust into the machine interior, hardness block, ball head after using coated with anti-rust oil, prevent rust;
Don’t disassemble the instrument without authorization, maintenance related matters please contact MITECH after-sale service department, 4000600280
Test force | 98.07N,tolerance ±2.0%10gf (0.098N)、25gf (0.245N)、50gf (0.49N)、100gf (0.98N)、 200gf (1.96N)、300gf ( |
Measuring range | 5~2500HV |
Convert ruler | HRA、HRB、HRC、HRD、HR15N、HR30N、HR45N、HR15T、HR30T、HR45T、HV、HK、HBW |
Display attributes | LED Digital Display |
Applying way of test force | Automatic (loading, holding, unloading) |
Conversion way of indenter | Manual |
Magnification of measure microscope | 100X(observing )400X(measuring) |
Test force holding time | 0~60s |
Minimum resolution | 0.0625μm |
Maximum height of specimen | 75mm |
Maximum width of specimen | 95mm |
Power supply | AC220V/50Hz |
Dimension | 441*270*480mm |
Main unit weight | 31kg |
如您在和我公司的交往中遇到問(wèn)題,可通過(guò)以下方式和我們?nèi)〉寐?lián)系!
您也可以撥打我們客戶服務(wù)熱線:400-0600-280
86-10-58859527
008613810243516
E-506, 1# of ShangDi East Road, Haidian District, Beijing, China
mvip@mitech-ndt.com